WARNING
Any replacement work carried out without the wheelchair manufacturer's permission will invalidate the control system's warranty.
PGDT accepts no liability for losses of any kind if the procedure and safety guidelines are not followed. These operations should
only be carried out by a trained Healthcare Technician.
To replace the Joystick Cable in the VR2 Joystick Module please follow the below procedure.
REMOVAL
- Isolate the Joystick Module by disconnecting the Joystick Cable from the Power Module.
- Remove the 5 retaining screws from the underside of the Joystick module.
- Disconnect the Serial Cable from the PCB.
- Remove the 2 retaining screws holding the Charger Socket and Cable.
- Slide the Charger Socket and Cable out of the plastic case.
REPLACEMENT
- Insert the new Charger Socket and Cable into the plastic case.
- Attach the Serial Cable from the new Charger Socket and Cable onto the PCB
- Secure the Charger Socket and Cable with the 2 retaining screws.
Tightening to 1.0Nm, 8.85 inch/lbs
- Reassemble the joystick module, ensuring the rubber gasket is correctly positioned, and replace the 5 retaining screws. Tightening to 1.0Nm, 8.85 inch/lbs
Incorrect fitting of the Gasket could seriously affect the Joystick Module’s resistance to moisture ingress.
- Connect the Joystick Cable to the Power Module.
- Turn the VR2 on.
If the replacement sequence has been unsuccessful the TruCharge display will flash 7 bars or the.
- Check all connections.
- Repeat the procedure.
- Should the procedure fail twice, try a new cable.
|