WARNING
Any replacement work carried out without the wheelchair manufacturer's permission will invalidate the control system's warranty.
PGDT accepts no liability for losses of any kind if the procedure and safety guidelines are not followed. These operations should
only be carried out by a trained Healthcare Technician.
To replace the Keypad of the VR2 Joystick Module please follow the below procedure.
- Disconnect Joystick Cable from the Power Module.
- Gently lift the top corner of the keypad, with a scalpel.
- Remove the damaged keypad.
- Ensure surface area is clear and clean.
- Place the new keypad into position, ensuring there are no gaps around the edges , and press firmly to stick.
- Turn control system on the and check the operation of the buttons.
Incorrect fitting of the Keypad could seriously affect the Joystick Module’s resistance to moisture ingress.
If trouble is experienced activating any of the buttons ensure that the pad is correctly positioned and firmly adhered to the joystick module's surface. |